Description:
The BD FACSymphonyâ„¢ system is a novel cell analyzer that leverages the inherent benefits of flow cytometry and enables the simultaneous measurement of up to 50 different characteristics of a single cell. This high parameter flow cytometer is a powerful analytical tool that enables scientists to identify and analyze distinctive phenotypes in heterogeneous populations.
Specifications | |
---|---|
Sorter/ Analyzer |
Analyzers |
Supplier |
BD Biosciences |
Catalog # |
Varied |
Dimensions |
83.8 x 76.2 x 80cm |
RUO/Clinical |
RUO |
Colors (up to) |
28 |
Lasers (up to) |
10 |
Laser Options |
355/375/405/420/445/458/460/473/488/505/514/532/552/561/568/588/592/628/637/640/674/660/685/730/785/980nm |
Customizable |
Yes |
Automation |
Optional |
Max Sample # |
384 |
Sample Formats |
1.5/2.0/5.0-mL tubes; optional high throughput sampler to acquire samples from a 96- or 384-well microtiter plate. |
Sample Cooling |
No |
Events/Sec Max |
40,000 |
Flow Rate |
Adjustable |
Measuring/Sort Time |
Inquire |
Detectors |
30 |
Sheath Bottle |
10 L |
Imaging |
No |
Microfluidic |
No |
Flat-top beam |
No |
Camera |
Cuvette |
Colinear |
No |
Instrument Configuration
LASER 355 |
FILTERS
379/28 |
470/15 450LP |
515/30 495LP |
586/15 570LP |
610/20 600LP |
660/20 640LP |
730/45 690LP |
820/60 770LP |
LASER 405 |
FILTERS
431/28 |
515/20 505LP |
586/15 570LP |
610/20 595LP |
670/30 640LP |
710/50 685LP |
750/30 735LP |
800/30 770LP |
LASER 488 |
FILTERS
515/20 505LP |
610/20 595LP |
710/50 685LP |
780/60 750LP |
||||
LASER 561 |
FILTERS
586/15 |
610/20 595LP |
660/40 640LP |
710/50 690LP |
780/60 750LP |
|||
LASER 637 |
FILTERS
660/20 |
730/45 690LP |
780/60 750LP |
|||||