Spark Red 718
Spark Red 718 (Ex-Max 695 nm/Em-Max 710 nm)
Spark Red™ 718 is a brighter alternative to Alexa Fluor® 700, adding to the fluorophore options for the red laser that can detect antigens with lower expression levels. It can be used on a conventional cytometer with a 633 nm red laser and 720/45 filter set (or equivalent filter used to detect Alexa Fluor® 700) spectral cytometers with appropriate instrument settings. On conventional cytometers, Spark Red™ 718 can be utilized with both APC and APC/Cyanine7. On spectral cytometers, it can be unmixed from APC, Spark NIR™ 685, Alexa Fluor® 660, APC/Cyanine7, and APC/Fire™ 810. As spectral properties of Spark Red™ 718 and Alexa Fluor® 700 are nearly identical, using them together in a multicolor panel is not recommended. With a medium brightness, Spark Red™ 718 is best suited for detecting antigens with low to moderate expression levels.
Excitation & emission spectra
Key facts
- Excitation maximum
- 695 nm
- Emission maximum
- 710 nm
- Optimal laser line
- 640
- Optimal filter
- 695/40
- Relative brightness
- 3 of 5
- Commercial antibodies
- 448 from 2 suppliers
Spark Red 718 has an excitation maximum at 695 nm and an emission maximum at 710 nm, a Stokes shift of 15 nm. It is optimally excited by the 640 laser line.
Frequently asked questions
- What is Spark Red 718?
- Spark Red™ 718 is a brighter alternative to Alexa Fluor® 700, adding to the fluorophore options for the red laser that can detect antigens with lower expression levels. It can be used on a conventional cytometer with a 633 nm red laser and 720/45 filter set (or equivalent filter used to detect Alexa Fluor® 700) spectral cytometers with appropriate instrument settings. On conventional cytometers, Spark Red™ 718 can be utilized with both APC and APC/Cyanine7. On spectral cytometers, it can be unmixed from APC, Spark NIR™ 685, Alexa Fluor® 660, APC/Cyanine7, and APC/Fire™ 810. As spectral properties of Spark Red™ 718 and Alexa Fluor® 700 are nearly identical, using them together in a multicolor panel is not recommended. With a medium brightness, Spark Red™ 718 is best suited for detecting antigens with low to moderate expression levels.
- What laser excites Spark Red 718?
- Spark Red 718 is optimally excited by the 640 laser line. Its excitation maximum is 695 nm.
- What filter should I use to detect Spark Red 718?
- A 695/40 filter captures the peak of the Spark Red 718 emission near 710 nm with minimal spillover.
- How bright is Spark Red 718?
- Spark Red 718 has a relative brightness of 3 out of 5.
- How many antibodies are available conjugated to Spark Red 718?
- FluoroFinder tracks 448 commercially available Spark Red 718 conjugates, searchable by target, clone, host and application.